Abbreviations: NCS = nerve conduction study; S-NCS = short-segment NCS; UNE = ulnar neuropathy at the elbow.
NCS with short-segment interval stimulations has been commonly used to increase the sensitivity for very focal lesions, because the effect of the abnormally slow segment is progressively diluted as more and more normal segments are included in the examination.5 A short-segment NCS (S-NCS) can be performed by stimulating the nerve at 2-cm intervals across the elbow in the flexed position.6 A significant conduction delay or focal change in amplitude or waveform morphology across a segment is generally considered abnormal; however, ulnar nerve displacement by elbow flexion may cause a more significant technical error in S-NCS than in routine NCS at the across-elbow segment.
In the present study, we investigated the effects of ulnar nerve displacement on S-NCS at the across-elbow segment. We also measured the angle of the elbow when the ulnar nerve was dislocated by elbow flexion.
| METHODS |
|---|
|
|
|---|
Real-time ultrasonographic study. A 7.5 to 12 MHz linear transducer (5000 System HDI; Philips Medical Systems NA, WA) was applied on the line between the medial epicondyle and olecranon of subjects lying supine on an examining table. We used the B mode to view the ulnar nerve in the postcondylar groove between the medial epicondyle and olecranon and used the color Doppler mode to differentiate arteries. A baseline nerve location was determined at zero degrees of elbow flexion, and we observed whether the ulnar nerve was displaced at any angle during elbow flexion using real-time ultrasonography. We classified the position of the ulnar nerve (during elbow flexion) into one of three types according to the classification system of a previous report.7 Type N (no dislocation) classification was assigned when the nerve moved anteriorly, but not to the tip of the epicondyle, during elbow flexion. The ulnar nerve position was classified as type S (subluxation) when the nerve moved onto the tip of the epicondyle. The position of the ulnar nerve was classified as type D (dislocation) when the nerve moved anteriorly beyond the tip of the epicondyle. In cases of type D, we measured the angle of the elbow using a goniometer when the ulnar nerve was dislocated (figure 1). To measure the angle, an axis of the goniometer was located on the skin of the medial epicondyle tip, and a proximal fixed arm of the goniometer was aligned parallel to the long axis of the upper arm. A distal movable arm was then attached to the medial surface of the forearm long axis. The elbow angle was defined as the angle made by the movement of the distal movable arm with forearm flexion from the extended position.
|
Short-segment nerve conduction study. S-NCSs for the ulnar nerve were performed in subjects with type S and D classifications by an electromyographer who was blind to the classification type. The S-NCSs were performed at an elbow position of approximately 135°. Prior to the S-NCS, we confirmed that none of the subjects among those with type S and D classifications had weakness in the ulnar innervated muscles or sensory changes in the fourth and fifth fingers. To achieve approximately 135° of elbow flexion in the most comfortable position possible, subjects were positioned supine on an examining table and asked to abduct their upper arm to 90° and to rest it on a pillow such that the hand approached the ipsilateral ear, resulting in elbow flexion. The midpoint of the ulnar groove between the tip of the medial epicondyle and the olecranon was marked (point P; figure 2. From this point, marks were placed along the course of the ulnar nerve at 2-cm intervals from 6 cm proximal to 4 cm distal.
|
The ulnar nerve was stimulated with abductor digiti minimi recording at each mark. The recording electrode was placed over the belly of the abductor digiti minimi, and the reference electrode was placed over the distal tendon. A constant current stimulator (2-cm intervals between cathode and anode) and a disposable, flat-surface electrode 1 cm in diameter were used for recording. Equipment used for the study was manufactured by Nicolet Biomedical (Viking IV, Madison, WI). A sensitivity setting of 5 mV/division, a sweep speed of 2 msec/division, and a filter setting of 5 Hz to 5 KHz was maintained throughout all measurements. The skin temperature of the subjects during the study was also maintained at or above 32 °C. Before the measurements were performed, we ensured that supramaximal stimulation was achieved and that adequate pressure was applied to the stimulating electrodes in order to enable focal stimulation without spread.
Latencies were measured from the stimulus to CMAP onset, and the amplitude measurements were calculated from baseline to negative peak. A conduction delay of greater than 0.5 msec or conduction block with an amplitude reduction of 50% or more across 2-cm segments were defined as abnormal.8
After the S-NCS, the nerves that were displaced by elbow flexion were tracked using ultrasonography and marked as ragged dots when elbow flexion was at 135°. The closest point from the tip of the medial epicondyle was marked as point P'. Marks were then placed along the course of the tracked ulnar nerve at 2-cm intervals from 6 cm proximal to 4 cm distal to point P' (ragged dots in figure 2). The S-NCSs were performed again by stimulating the ragged dots.
| RESULTS |
|---|
|
|
|---|
Short-segment nerve conduction study. All 13 cases of ulnar nerve dislocation at the elbow exhibited delayed latency (12 subjects; mean 0.62 msec, range 0.6 to 0.8 msec) or a conduction block around the medial epicondyle during the S-NCS with black dot stimulation. Of the 13 subjects with conduction block, the amplitude drop points were P in 6 subjects and P2 in 3 subjects; the mean amplitude drop rate in these 9 subjects was 68.9% (range 58.8 to 92.2%). The other 4 subjects showed complete conduction block (no evoked compound muscle action potential) at point P2. One of the 4 subjects with complete conduction block showed an amplitude drop of 34% at point P. However, no abnormal findings were detected in the elbows of the subjects with ulnar nerve subluxation during elbow flexion.
The abnormal findings in the 13 elbows with ulnar nerve dislocation were transformed to normal waveforms and latencies using ragged dot stimulation (figure 3).
|
| DISCUSSION |
|---|
|
|
|---|
UNE is the second most common upper extremity compression neuropathy10; however, the diagnosis of UNE is not straightforward. Most physicians suspect UNE when a patient complains of occasional ulnar distribution numbness and paresthesias; however, these symptoms could be caused by a variety of conditions, such as lower cervical radiculopathy, lower trunk involvement of the brachial plexus, Guyon canal palsy, and even myofascial pain syndrome. The diagnostic yield of electrodiagnosis of UNE is low, and the interpretation of the data is often quite difficult. In our previous study, we suggested the possibility of false negatives in the detection of UNE if a patient has ulnar nerve dislocation during elbow flexion.1 If a patient is seen who is suspected of having UNE, but nonetheless shows normal findings in routine NCS, S-NCS could be used as one of the next steps to increase sensitivity of the diagnosis. Indeed, the results of the present study showed that the S-NCS performed with elbow flexion in patients with ulnar nerve dislocation at the elbow could lead to a significant diagnostic error. All elbows showed conduction block or significant latency delays in the ulnar nerve at the across-elbow segment, which is important, because the identification of these signs are important criteria for the diagnosis of UNE.6 Therefore, significant caution is needed when using S-NCS. Although elbow flexion at 90 or 135° is recommended during NCS to straighten the ulnar nerve in the ulnar groove, the technical error induced by ulnar nerve dislocation during elbow flexion should not be overlooked. In the present study, most elbow angles when the ulnar nerves were dislocated were more than 90°. Only three elbows were at an angle less than 90°, and two of the three elbows were very close to 90°, both at an angle of 87°. This finding indicates that there is less chance of ulnar nerve dislocation during NCS at the across-elbow segment when the degree of elbow flexion is less than 90°.
To our knowledge, although there have been no studies that have reported that dislocation of the ulnar nerve predisposes a patient to ulnar neuropathy, recurring ulnar nerve dislocation at the elbow may cause bothersome symptoms in some subjects via repeated irritation or enhanced exposure to injury. Therefore, detection of ulnar nerve dislocation using ultrasonography would be helpful in evaluating patients with normal conduction velocities in routine nerve conduction study at the across-elbow segment, despite clinical symptoms of highly suspected UNE. If ultrasonography is not available in a clinic, an alternative study could include S-NCS, which should be performed with two elbow positions, with one at less than 90 degrees and the other at full flexion. The appearance of conduction delay or block with full flexion would support either dislocation or transient compressive neurapraxia, either of which might be important in a symptomatic patient after other causes of symptoms have been excluded. Future studies are needed to investigate the reliability of this potential approach and the importance of dislocation as a predisposing factor for UNE.
S-NCS for assessment of the ulnar nerve at the across-elbow segment requires special caution due to the possibility of technical error caused by elbow position. Many laboratories use a variety of elbow angles in order to minimize measurement errors due to an ulnar nerve redundancy; however, such practices may overlook another source of artifact, especially since we observed that ulnar nerve dislocation induced by elbow overflexion could lead to significant technical error. The results of our study suggest that S-NCS at the elbow should be performed with the elbow flexed at less than 90 degrees. Future studies are needed to investigate the ideal range of the elbow angle during S-NCS in order to avoid technical error caused by the position of the elbow.